Your Advantages
- Regular maintenance protects and maintains your investment
- Maintenance carried out in accordance with KaVo quality standards
- Remain up to date and compliant with hygiene and maintenance standards
- Reduces downtime and unplanned repair costs
Functional safety | Safety of the dentists, dental practice team and patients | Legally required | |
---|---|---|---|
Testing of all important functions, e.g. hygiene functions, safety switch, etc. | | | |
Cleaning the suction system, Aquamat and selection unit | Replacing wearing parts | | | |
Replacing worn parts in the water block | Inspecting for leaks | | | |
Inspecting the O-rings on the doctor’s and assistant’s element, hoses and pressures | Replacing worn parts | | | |
Check Amalgam separator | | | |
Safety inspection in accordance with IEC (DIN EN) 62353 | | | |
Your unit will perform many demanding services for you over the course of its service life.
treatments and
operating hours
will be performed in a treatment unit over an average of
years.
litres of water
are used during this time.
Regular maintenance is therefore essential.
Sequence
All maintenance steps follow KaVo's maintenance schedule and are documented step by step. You will receive a maintenance log as evidence
1. Selective plunger
- Check function, clean, and replace O rings
- Disassemble and clean selective housing, replace strainers
- Disassemble selective plunger, replace springs (E70/E80) and seals
2. Aquamat
- Check overall function
- Replace reservoir and clean top
3. Suction flow valve / external suction
- Replace O ring and piston packing
4. Water unit
Check overall function / check for leakage
- Replace water filter, membrane, and seals
- Replace non-return valve
- Replace plug-in filter, water inlet unit fitting and O rings
- Check system water pressure and adjust if required
5. Patient unit
- Tumbler spout: check operation and seal
- Replace jet control and O ring
- Bowl flush: check operation and seal
- Replace spittoon bowl strainer
6. Dentist element
- Replace return air filter
- Disassemble and clean control valve
- Replace plunger, O-ring, and membrane